n. b. This is not a foolproof guide. It is very important that one has to know what he is doing, why he is doing it, and what’s happening in Mr. Freeze if he does it.
STOPPING THE CIRCULATION 1. If the BOOSTER PUMP is being used, do the following: a. Switch off the QSB valve. b. Close valve #13. c. Turn off the BOOSTER PUMP. 2. If only the 3He ROTARY PUMP is in use, do the following: a. Close #12. b. Turn off the ROTARY PUMP. 3. Open valve #2. 4. Close the C needle valve on top of the refrigerator. 5. Turn off the 4He ROTARY PUMP. 6. Open #31.
RETURNING He MIXTURE TO THE DUMPS The best time to do this is when the temperature reaches 4.2 K.
1. Close 8. 2. Open 1. 3. 3 should be open. 4. Close 5. 5. Open 9. 6. Turn on the He3 ROT PUMP. 7. Slowly open 12 until it is fully open. 8. Make sure dumps valve are open. 9. When 12 is fully open, wait a few minutes. 10. Close 1, 7, and 9 in that order. 11. Turn off the He3 ROT PUMP. 12. Close dumps valve.
CLEANING/PUMPING NITROGEN TRAP